首页|The Application of ANOVA in Study of Fluorescence Detection System for Petroleum Substence in Water

The Application of ANOVA in Study of Fluorescence Detection System for Petroleum Substence in Water

扫码查看
According to the spectral characteristic that petroleum substance would emit fluorescence as they are excited by light of specifically wavelength, a system for measuring petroleum concentration in water based on fluorescence analysis was designed。 It adopted a pulsate Xenon lamp as an excitation light source, used a probe to realize direct measurement and used optic fiber to transmit excitation light and fluorescence。 A charge couple device was employed to detect the spectra and accomplish photoelectric conversion。 Analysis of variance (ANOVA) and F-test was adopted to check the work curves for the ability of predict concentration。 The experiment results showed that the system can realize the measurement of petroleum substance concentration in water quite well。

petroleumfluorescenceoptic fiberANOVAF-test

Pan Zhao、Wang Yutian、Yang Zhonghui

展开 >

Key Laboratory of Measurement Technology and Instrumentation of Hebei Province, Yanshan University Qinhuangdao, China

International conference on intelligent information technology application;IITA2010

Qinhuangdao(CN);Qinhuangdao(CN)

2010 4th International conference on intelligent information technology application.;vol. 2.

p.257-260

2010