Influence and correction of self-absorption of cylindrical samples on HPGe γ-spectrum measurement results
Objective To analyze the influence of the self-absorption effect of different density samples on the measurement results of HPGe γ-spectrometer,and discuss the correction method of the cylindrical sample self-absorption in the measurement of HPGe γ-spectrometer.Methods Ten kinds of samples with different densities were measured by HPGe y-spectrometer,and the national standard material of thorium powder and uranium-radium balance powder were added in the samples.The γ-ray energy points of 232Th and 226Ra in the sample spectrum were analyzed,and the relative self-absorption factor(F)was calculated according to the activity formula.The correction curve was fitted with the density of cylindrical samples and the relative self-absorption coefficient,and the self-absorption coefficient was calculated by the mass attenuation coefficient of γ-ray of a known substance for verifying the accuracy of the self-absorption curve that fitted before.Results There were no statistical differences(P>0.05)between the mean values of the energy points of y-rays count rate in the samples added thorium of 238.6 keV(t=-2.128),338.7 keV(t=-2.002),583.1 keV(t=-2.086),911.2 keV(t=-2.165)and that in the samples added aluminum powder of 295.2 keV(t=-1.541)、351.9 keV(t=-1.557)、609.3 keV(t=-1.498)and 1 120.3 keV(t=-1.688).The differences were not statistically significant.When the density of cylindrical sample is different from that of calibration source,the relative self-absorption coefficient deviation was greater with the density difference.The relative self-absorption coefficient decreased with the increase of γ-ray energy.All deviations of self-absorption coefficient of energy points calculated by formula and that measured by experiment were less than 10%.Conclusions If the difference exists between the density of the sample and the calibration source,the self-absorption correction should be carried out.If the densities are same,the effect of self-absorption is greater in the low energy region than that in the high energy region.The result got by experimental measurement is more reliable than that calculated by formula.The self-absorption correction curve of the cylindrical sample fitted in this experiment can be used for daily measurement correction in the laboratory.