GB/T 42125.1-202X "Safety requirements for electrical equipment for measurement,control,and laboratory use-Part 1:General requirements"(IEC 61010-1(ED 3.0),IDT) will soon be released and implemented,and the differences in routine tests brought about by the standard change will have a far-reaching impact on the measurement and control instrument industry.Based on annex F of the standard,the authors interpret and analyze the original intention,test steps and precautions of the ground continuity and dielectric strength test items in routine tests for readers,aiming at helping industry manufacturers to adjust countermeasures as soon as possible to cope with the differences in the standard version.
关键词
测量、控制和实验室用电气设备/例行试验/接地连续性/介电强度/IEC/61010-1
Key words
electrical equipment for measurement,control,and laboratory use/routine test/grounding continuity/dielectric strength/IEC 61010-1