Determination of Arsenic in Coal by Atomic Fluorescence Spectrometry
This paper mainly studied the atomic fluorescence spectrometry determination of arsenic in coal, the method of recovery of the method (95.75~98.34)%, RSD<2.36%.With arsenic molybdenum blue spectrophotometric method determination of arsenic in coal, and discusses the accuracy and precision of this method, the results show that the method is simple, rapid, high sensitivity, the determination of arsenic concentration in large range, is suitable for the determination of arsenic in coal.