首页|Measurement of thin liquid film thickness in pipes based on optical interferometry

Measurement of thin liquid film thickness in pipes based on optical interferometry

扫码查看
Accurate measurement of the thin liquid film thickness in pipes is the foundation for studying the characteristics of the film.In this paper,an interferometry-based measurement of liquid film thickness in transparent pipes is developed,which can greatly improve the accuracy,extend the lower limit of measurement and provide a new technical approach for the calibration and traceability.The light intensity distribution is established based on the optical path analysis and a mathematical model.A new algorithm to solve the direction ambiguity is developed to reconstruct the phase distri-bution.Besides,the effect of the pipe wall is taken into account,which can be suppressed by image subtraction and enhancement technology.The proposed method is of high accuracy and robustness,whose reconstruction errors are 0.064%and 0.25%for the smooth and slight fluctuating liquid films,respectively.

XUE Ting、WU Yan

展开 >

School of Electrical and Information Engineering,Tianjin University,Tianjin 300072,China

国家自然科学基金国家自然科学基金

6207132561828106

2022

光电子快报(英文版)
天津理工大学

光电子快报(英文版)

EI
影响因子:0.641
ISSN:1673-1905
年,卷(期):2022.18(8)
  • 1