During the robot-assisted pelvic fracture reduction,the needle-tissue interactive deformation characteristic is not clear,which affects the accuracy of robotic surgery.In this paper,a layered rig-id-flexible coupling model is proposed,and the needle-tissue interactive deformation under the load-ing is analyzed by the Rayleigh-Ritz method,in accordance with the principle of minimum potential energy.The pelvic musculoskeletal tissue is reversely reconstructed,and the structure of the bone is segmented into cancellous bone and cortical bone.The Mooney-Rivlin five-parameter hyperelastic model is used to simulate muscle,and the Ogden hyperelastic model is used to simulate adipose tis-sue.Finite element simulation is performed by loading different magnitudes of forces.The accuracy of the rigid-flexible coupling model is 0.432 mm,which indicates the correctness of the needle-tis-sue interactive deformation theory analysis.

Key words

pelvic fracture reduction/rigid-flexible coupling/needle-tissue interactive de-formation

引用本文复制引用

出版年

2024
高技术通讯(英文版)
中国科学技术信息研究所(ISTIC)

高技术通讯(英文版)

影响因子:0.058
ISSN:1006-6748
段落导航相关论文