首页|生物安全柜质量检测仪雾化量校准方法研究

生物安全柜质量检测仪雾化量校准方法研究

扫码查看
针对生物安全柜质量检测仪易出现雾化量不足及其校准方法的缺失,研究其雾化量校准方法.基于搭建的雾化量校准装置,通过提出雾化量校准方法,确定其技术指标,结合校准案例进行验证以及对相关影响因素进行分析.依据校准方法对生物安全柜质量检测仪的雾化量示值误差(质量密度法和时间流量法)的验证结果符合技术要求.成功建立生物安全柜质量检测仪雾化量校准方法,可为《生物安全柜质量检测仪校准规范》的编写提供有益参考,促进生物安全柜质量检测仪的量值溯源和标准化.
Research on atomization amount calibration method of biosafety cabinet quality testers
Aiming at the lack of atomization amount and the lack of calibration method of biosafety cabi-net quality detector,the atomization amount calibration method was studied.Based on the atomization a-mount calibration device,the atomization amount calibration method is proposed,the technical indica-tors are determined,and the calibration case is verified and the relevant influencing factors are ana-lyzed.According to the calibration method,the verification results of the atomization quantity indication error(mass density method and time flow method)of the biosafety cabinet quality tester meet the tech-nical requirements.The successful establishment of the atomization calibration method of biosafety cabinet quality detector can provide a useful reference for the preparation of the"Calibration Specification for BioSafety Cabinet Quality Tester",and promote the traceability and standardization of the measurement value of biosafety cabinet quality tester.

biosafety cabinet quality testeratomization amountcalibrationmass density methodtime flow method

司启益、张新白、崔宏恩

展开 >

江苏省计量科学研究院,江苏南京 210023

生物安全柜质量检测仪 雾化量 校准 质量密度法 时间流量法

江苏省市场监督管理局科技项目

KJ204112

2024

工业计量
冶金自动化研究设计院 中国计量协会

工业计量

影响因子:0.256
ISSN:1002-1183
年,卷(期):2024.34(1)
  • 13