Currently,radiation therapy health centers and the development of medical devices have be-come important technical backbone in the world medical industry.With the deepening development of modem Internet emerging technology and radiation physics technology,radiotherapy technology has made great progress.However,there are still many urgent problems to be solved in the current clinical practice of radiation therapy technology.In order to promote precise positioning and effective treatment of radiation therapy,this article is based on the development of central detection devices such as radiation therapy simulation positioning machines.It involves a measuring device for three parameters of radiation therapy simulation positioning machines in the medical and health industry:"equal center indication accuracy","defining the offset between the radiation beam axis and the equal center",and"consistency between the light field and the radiation field".This instrument belongs to the field of radiation medical devices and is used for quality inspection of the application of radiation therapy simulation positioning ma-chines.Through research and design,a radiation therapy simulation positioning machine and other cen-tral detection device have been developed.After experimental verification,the device can meet the tech-nical requirements of the national metrological verification regulation JJG 1028-2007"Radiation Therapy Simulation Positioning X-ray Radiation Source".In order to promote radiation therapy,Continuous pro-gress provides corresponding technical support.
radiotherapysimulated positioning machinewaiting center detection devicemedical ap-paratus and instruments