吹扫捕集/气相色谱-质谱联用仪测定饮用水水源地中的1,3-丁二烯和1-溴丙烷

扫码查看
使用吹扫捕集/气相色谱-质谱联用仪,采用SIM扫描模式对饮用水水源地中的1,3-丁二烯、1-溴丙烷进行了测定,并进行了方法学考察.结果表明,2种物质的线性范围良好,1,3-丁二烯、1-溴丙烷的相关系数分别为0.999 9、0.998 0;检出限分别为0.2和0.3 μg/L,定量限分别为0.8和1.2 μg/L;高、中、低3个浓度的相对标准偏差在2.6%~7.6%之间;高、中、低3个浓度实际样品的加标回收率在84.0%~122%范围内;替代物的加标回收率在70.4%~103%之间.该方法在本实验室具有较低的检出限、良好的线性和精密度、较高的正确度.
Determination of 1,3-butadiene and 1-bromopropane in drinking water source by purge-trap/gas chromatography-mass spectrometry
1,3-butadiene and 1-bromopropane in drinking water source were determinated by purge-trap/gas chromatography-mass spectrometry using the mode of SIM,and the methodology of this method was investi-gated.The results showed that,the linear range of the two pollutants was good,the correlation coefficients of 1,3-butadiene and 1-bromopropane were 0.999 9 and 0.998 0 respectively;the detection limits were 0.2 and 0.3 μg/L respectively,and the quantitative limits were 0.8 and 1.2 μg/L respectively.The relative standard deviations(RSD)of accuracy of high,middle and low concentrations were between 2.6%to 7.6%;and the standard recovery rates of the actual samples with high,middle and low concentrations ranged from 84%to 122%;while the stan-dard recovery rates of the substitutes were 70.4%-103%.It showed that,the method has low detection limits,good linearity and precision,and high accuracy.

purge-trap/gas chromatography-mas spectrometry1,3-butadiene1-bromopropanedrinking water sourcesprecisionstandard recovery rate

董凤娟、黄现统、张凤菊、许昶磊、黄平

展开 >

山东省枣庄生态环境监测中心,山东 枣庄 277000

山东省生态环境监测中心,济南 250101

枣庄市环境保护科学研究所有限公司,山东 枣庄 277000

吹扫捕集 气相色谱-质谱联用仪 1,3-丁二烯 1-溴丙烷 饮用水水源地 精密度 加标回收率

2024

工业用水与废水
东化工程科技股份有限公司

工业用水与废水

CSTPCD
影响因子:0.649
ISSN:1009-2455
年,卷(期):2024.55(2)
  • 12