月球直射辐照度仪同光轴装调及传递定标

扫码查看
采用激光水平仪与平面反射镜相结合的方式,在实验室模拟月球对追踪相机与探测镜头进行同光轴装调,避免未同光轴导致测试不到信号或者信号太小造成数据不准确的问题,实验结果显示实验室同光轴装调误差在±0.03°以内.为保证仪器定标系数的准确性,采用已溯源于国家计量基准的标准灯传递定标微光灯的方式,利用激光测距仪与激光水平仪、平面反射镜相结合的方式保证传递定标精度,解决了直接使用标准灯定标带来的探测器非线性响应误差和长定标距离误差的难题.为地基对月观测仪器建立高精度的月球辐射模型提供了一种思路与解决方法.
Co-axis Alignment and Radiance Transfer Calibration of Direct Lunar Irradiance Meter
With the continuous development of space remote sensing instruments and the further improvement of the quantitative application requirements of remote sensing products,the necessity and importance of high-precision calibration of space remote sensing instruments are increasingly apparent.The moon's reflectance changes by 10-8/year,and its high stability in reflectance,as well as the repeatability of radiance at the same observation geometry angle,make the moon very suitable as a reference radiation source and a hot research topic in the field of radiation calibration at home and abroad.However,the key to achieving this function is the need for a large amount of ground-based observation data to support the establishment of a high-precision lunar radiation model.Therefore,it is urgent to develop radiation instruments that have long-term precise tracking and measurement capabilities on the moon.Accurate test data and laboratory radiation calibration are two key factors that directly affect the acquisition of high-precision ground-based observation data for ground-based lunar radiation instruments.Among them,the tracking camera and signal detection lens being on the same optical axis are the key factors affecting the accuracy of the test,and the low-light radiation calibration coefficient is the key factor affecting the calibration accuracy.The tracking camera's full field of view of the moon's direct irradiance meter developed in this paper is 2.13°×1.60°,and the signal detection lens's full field of view is 1°.Because both fields of view are small,it is necessary to strictly parallel the optical axis of the tracking camera and the signal detection lens.If they are not adjusted to the same optical axis,it will cause the tracking camera to track the moon in real-time,the moon is at the center of the tracking camera's field of view but not in the center of the detection lens's field of view,resulting in no signal being detected or inaccurate data due to the signal being too weak.To ensure the accuracy of test data,this paper combines a laser level instrument with a flat reflection mirror to simulate the moon in the laboratory and adjust the tracking camera and detection lens on the same optical axis,avoiding inaccurate data caused by not being on the same optical axis.The experimental results show that the error of the same optical axis adjustment in the laboratory is within±0.03°.In addition,the visible,infrared 1,and infrared 2 modules transmit radiation calibration,determining the coefficient relationship between the detector's output Digital Numbers(DN)value and the irradiance,which also directly affects the establishment of the lunar radiation model.To ensure the accuracy of the instrument's calibration coefficient,this paper uses a standard lamp traced back to the national metrology benchmark to transmit the calibration of the low-light lamp,using a combination of a laser rangefinder,a laser level instrument,and a flat reflection mirror to ensure the accuracy of the transmission calibration and solve the problems of non-linear response errors of the detector and long calibration distance errors caused by direct use of the standard lamp.In theory,this can improve the accuracy of low-light calibration.This paper aims to provide a universal and feasible solution to solve the problem of limited same optical axis adjustment due to environmental conditions affecting external field adjustment and to solve the problem of transmission radiation calibration accuracy and detector non-linear response using low-light sources for the moon's direct irradiance meter in the laboratory.This solution provides a way of thinking and a solution for laboratory optical axis adjustment and transmission calibration of integrated automatic observation remote sensing instruments with tracking and testing capabilities,and has important reference value.

Radiation calibrationLunar direct illumination meterLaser level instrumentCo-optical axisTransfer calibration

王鑫蕊、李新、潘琰、李萍、张权、韦玮、秦萌萌

展开 >

中国科学院合肥物质科学研究院 安徽光学精密机械研究所,合肥 230031

中国科学技术大学,合肥 230026

辐射定标 月球直射辐照度仪 激光水平仪 同光轴 传递定标

国家重点研发计划国家重点研发计划国家自然科学基金中国科学院合肥研究院院长基金

2022YFB39032002022YFB390320142105139YZJJ202309?TS

2024

光子学报
中国光学学会 中国科学院西安光学精密机械研究所

光子学报

CSTPCD北大核心
影响因子:0.948
ISSN:1004-4213
年,卷(期):2024.53(4)
  • 16