核电子学与探测技术2024,Vol.44Issue(6) :1125-1132.

非电离室型探测器在深度剂量扫描中的应用研究

Research on the Application of Non-ionization Chamber Detector in Deep Dose Scanning

李毅华 吴晗 王志鹏 王坤 金孙均
核电子学与探测技术2024,Vol.44Issue(6) :1125-1132.

非电离室型探测器在深度剂量扫描中的应用研究

Research on the Application of Non-ionization Chamber Detector in Deep Dose Scanning

李毅华 1吴晗 2王志鹏 3王坤 3金孙均3
扫码查看

作者信息

  • 1. 安徽理工大学 机电工程学院,安徽 淮南 232001
  • 2. 安徽理工大学 机电工程学院,安徽 淮南 232001;中国计量科学研究院,北京 102200
  • 3. 中国计量科学研究院,北京 102200
  • 折叠

摘要

为研究非电离室型探测器在深度剂量分布扫描中的响应特性,通过采用平板电离室、半导体和宝石探测器对医用光子束和电子束深度剂量分布进行扫描,并以平板电离室作为参考,分析比较半导体和宝石探测器在最大剂量深度、射线质等参数来探究其在深度剂量上的性能.结果表明:医用光子束下除两支探测器外,其余探测器获得的射线质与参考值的最大相对偏差为-0.67%,符合TRS-398 报告高能光子束允许的不确定度 1.0%;电子束<10 MeV时,各能量下非电离室型探测器获得的射线质与参考值的平均相对偏差最小为 1.91%;能量≥10 MeV时,平均相对偏差最大为1.01%,除两支探测器外,其余探测器的相对偏差最大为-1.15%,符合TRS-398 报告高能电子束允许的不确定度 1.2%.P型半导体探测器获得的射线质与参考值的相对偏差在±0.85%内,比 E 型半导体探测器的相对偏差范围±0.67%大.结论:非电离室型探测器在高能光子束下影响较小;电子束<10 MeV 时影响较大,能量≥10 MeV时影响较小,且尽量选择E型半导体探测器,并考虑探测器外壳材质.

Abstract

In order to study the response characteristics of non-ionization chamber detectors in depth dose distribution scanning,the depth dose distribution of medical photon beam and electron beam was scanned by using plane parallel ionization chamber,semiconductor and diamond detectors.The plane parallel ionization chamber is used as a reference to analyze and compare the parameters of semiconductor and gem detectors in maximum dose depth and radiation quality to explore their performance in depth dose.Results showed:under the medical photon beams,except for two detectors,the maximum relative deviation between the radiation quality obtained by the other detectors and the reference value is -0.67%,conforms to TRS-398 report high-energy photon beam allowable uncertainty of 1.0%;when the electron beam is less than 10 MeV the minimum average relative deviation between the radiation quality obtained by the non-ionization chamber detector and the reference value at each energy is 1.91%;when the energy is≥10 MeV,the maximum average relative deviation is 1.01%,except for two detectors,the maximum relative deviation of the other detectors is -1.15%,conforms to TRS-398 report high-energy electron beam allowable uncertainty of 1.2%.The relative deviation between the measured radiation quality and the reference value of the P-type semiconductor detector is within±0.85%,which is larger than the relative deviation range of±0.67%of the E-type semiconductor detector.Conclusion:The non-ionization chamber detector has little effect under high-energy photon beams.When the electron beam is less than 10 MeV,the influence is greater,and when the energy is≥10 MeV,the influence is smaller.The E-type semiconductor detector is selected as far as possible,and the material of the detector shell is considered.

关键词

电离室/半导体探测器/深度剂量/射线质/医用光子束和电子束

Key words

ionization chamber/semiconductor detector/depth dose/radiation quality/medical photon beam and electron beam