Principle of Automatic Analyzer for Fluorine and Chlorine in Coal and Analysis of Its Advantages and Disadvantages in Experiments
Accurate instrument detection is called the development direction of laboratory detection,especially in the detection of toxic and harmful components,replacing manual by equipment is an inevitable development trend.Sande FCl3000 tester combines high-temperature hydrolysis and potentiometric titration to achieve full automation,replacing manual testing in the laboratory and testing in accordance with national standards.This paper introduces the principle of the experimental equipment and compares the performance of the equipment in the work with the manual operation of the method specified in the national standard,including the scope of application,experimental environment,experimental time,operation safety,accuracy of results and so on.The equipment is inferior to manual operation in operating time and result accuracy,and the equipment is superior to manual operation in experimental environment and safety as well as coherence.Through the above research,operators can know more about the equipment to get accurate and reliable experimental results.
automatic analyzer for fluorine and chlorinehigh temperature hydrolysispotentiometric titration