Determination of 65 Volatile Organic Compounds in Environmental Soil
According to the"Determination of Volatile Organic Compounds in Soil and Sediments by Purge and Trap/Gas Chromatography-Mass Spectrometry HJ 605-2011",65 VOCs in environmental soil were detected by gas chromatography-mass spectrometry as a pre-treatment step.The internal standard method was used for quantification.The detection limit of 65 VOCs was 0.2~1.7ng/g,the relative standard deviation(RSD)was 0.04%~48.71%,and the recovery rate was 70.5%~112.3%.Comply with the requirements of relevant technical documents such as HJ 605-2011 and the Environmental Monitoring Specification for the determination of VOCs in environmental soil.
purge and capturevolatile organic compoundsgas chromatography-mass spectrometry