首页|全自动固相萃取-气质联用法测定废水中硝基苯类化合物

全自动固相萃取-气质联用法测定废水中硝基苯类化合物

扫码查看
采用全自动固相萃取处理废水,气相色谱-质谱联用法测定废水中的15种硝基苯类化合物,通过优化试验条件,使方法在0.100 mg/L~10.0 mg/L范围内线性良好.方法检出限为0.012 μg/L~0.038 μg/L,标准溶液11次测定结果的RSD为1.0%~7.5%.对实际水样做2个质量浓度水平的加标回收试验,平均加标回收率为78.0%~107%.
Determination of Nitrobenzene Compounds in Wastewater by Automatic Solid Phase Extraction-Gas Chromatography-Mass Spectrometry
Fifteen nitrobenzene compounds in wastewater were extracted by automatic solid phase and deter-mined by gas chromatography-mass spectrometry(GC-MS).By optimizing the test conditions,the method had good linearity in the range of 0.100 mg/L to 10.0 mg/L.The method detection limits were from 0.012 μg/L to 0.038 μg/L.The RSDs of 11 measurements of standard solution were from 1.0%to 7.5%.The average spiked recoveries were from 78.0%to 107%when adding two mass concentration levels to practical samples.

Nitrobenzene compoundsSolid phase extractionGC-MSIndustrial wastewater