Development and Performance Verification of a Portable Gas Automatic Sampling Device
According to the calibration requirements of flame ionization detector(FID)and thermal conductivity detector(TCD)in JJG 700-2016"Verification Regulations for Gas Chromatographs",a portable automatic gas sampling device is designed and developed.The device can perform continuous automatic gas injection in a volume range of(0.1~5.0)mL,and the seven consecutive peak area repeatability(RSD)measured by the performance verification experiment is 0.9%.At the same time,the stability of the device itself and the convenience of operation can effectively improve the measurement quality and efficiency of the gas chromatograph.
automatic gas injection devicegas chromatographperformance verification