计量与测试技术2024,Vol.51Issue(5) :24-27.DOI:10.15988/j.cnki.1004-6941.2024.5.007

水质臭氧分析仪校准系统介绍及溯源方法研究

Introduction to Calibration System of Water Quality Ozone Analyzer and Study on Traceability Method

李学辉 李亚飞 张亚飞 罗军 陈希尧
计量与测试技术2024,Vol.51Issue(5) :24-27.DOI:10.15988/j.cnki.1004-6941.2024.5.007

水质臭氧分析仪校准系统介绍及溯源方法研究

Introduction to Calibration System of Water Quality Ozone Analyzer and Study on Traceability Method

李学辉 1李亚飞 1张亚飞 1罗军 2陈希尧1
扫码查看

作者信息

  • 1. 上海市计量测试技术研究院
  • 2. 广州分析测试中心
  • 折叠

摘要

研制水质臭氧分析仪的校准系统,并对其进行不确定度评定,使用紫外分光光度计,对臭氧水的吸光度定值和摩尔吸光系数用碘量法进行量值溯源.结果表明:该系统稳定性为3%/10min,扩展不确定度为0.05mg/L,k=2,可用于该分析仪的校准.

Abstract

The calibration system of ozone water quality analyzer was developed and its uncertainty was evalua-ted.The absorbance and molar absorbance coefficient of ozone water were determined by ultraviolet spectrophotome-ter and traced by iodine method.The results show that the stability of the system is 3%/10min,the extended uncer-tainty is 0.05mg/L,k=2,and it can be used for the calibration of the analyzer.

关键词

水质臭氧分析仪/校准系统/不确定度/量值溯源

Key words

water quality ozone analyzer/calibration system/uncertainty/quantity traceability