首页|水质臭氧分析仪校准系统介绍及溯源方法研究

水质臭氧分析仪校准系统介绍及溯源方法研究

扫码查看
研制水质臭氧分析仪的校准系统,并对其进行不确定度评定,使用紫外分光光度计,对臭氧水的吸光度定值和摩尔吸光系数用碘量法进行量值溯源.结果表明:该系统稳定性为3%/10min,扩展不确定度为0.05mg/L,k=2,可用于该分析仪的校准.
Introduction to Calibration System of Water Quality Ozone Analyzer and Study on Traceability Method
The calibration system of ozone water quality analyzer was developed and its uncertainty was evalua-ted.The absorbance and molar absorbance coefficient of ozone water were determined by ultraviolet spectrophotome-ter and traced by iodine method.The results show that the stability of the system is 3%/10min,the extended uncer-tainty is 0.05mg/L,k=2,and it can be used for the calibration of the analyzer.

water quality ozone analyzercalibration systemuncertaintyquantity traceability

李学辉、李亚飞、张亚飞、罗军、陈希尧

展开 >

上海市计量测试技术研究院

广州分析测试中心

水质臭氧分析仪 校准系统 不确定度 量值溯源

2024

计量与测试技术
成都市计量监督检定测试所

计量与测试技术

影响因子:0.175
ISSN:1004-6941
年,卷(期):2024.51(5)