Quality Control Testing Practice and Improvement Measures of Defibrillation Monitor in Prefectures and Cities Public Tertiary Hospitals
Objective:The objective of this study was to discuss the quality control testing method for defibrillation monitors and en-hance their safety and effectiveness.Methods:FLUKE professional testing equipment was used to assess the appearance,power-up functionality,electrical safety,energy accuracy,charging time,synchronization delay time,and heart rate error of 45 active defibrilla-tion monitors in a certain municipal public hospital.Results:The overall pass rate for the 45 defibrillation monitors was found to be 73%.Specifically,the appearance qualified rate was 93%,boot qualified rate was 75%,energy accuracy qualified rate reached 97%,while electrical safety,charging time,synchronization delay time and heart rate error all achieved a perfect score of 100%.Based on identified causes for non-compliance issues during initial testing phase,improvement measures were implemented resulting in success-ful retesting outcomes.Conclusion:Quality control testing for defibrillation monitors can effectively mitigate usage-related safety risks and provide valuable guidance for medical equipment quality control and management within the region.
Prefecture-level Public HospitalsDefibrillation MonitorQuality ControlImprovement Measure