ICP-OES Determination of 7 elements in zirconium-yttrium alloy
An analytical method for the determination of Cr,Si,Y,Fe,Ca,Ti,and Hf elements in zirconium and yttrium alloys by inductively coupled plasma atomic emission spectrometry was established.Through the spectral interference experiment,the ana-lytical spectral lines of each element were determined:Y 360.07 nm,Cr 267.716 nm,Si 288.158 nm,Fe 238.204 nm,Ca 373.690 nm,Ti 336.121 nm,Hf 264.141 nm.Giving values of recovery and RSD's(n =7)in the ranges of 90%~110%and 0.11%~0.90%,respectively.