Study and Application of Determination of Acarbonyl Compounds in Ambient Air by High Performance Liquid Chromatography
A method for the determination of carbonyl compounds in ambient air by high performance liquid chromatography(HPLC)was established by using 2,4-dinitrophenylhydrazine(DNPH)coated silica gel filled tubes to collect carbonyl compounds in ambient air.The results showed that the test method had a good linear relationship,and the linear coefficient ranged from 0.9998 to 1.000;The detection limit of the method was 0.65~2.46 μg·m-3,the lower limit of determination was 2.61~9.85 μg·m-3;The relative standard deviations of low,medium and high mass concentrations were 0.9%~8.5%,0.5%~3.6%and 0.6%~2.8%respectively;The actual sample spiking recovery range was 89.8%~116%.The method has the advantages of good stability,simple operation and high sensitivity,and is suitable for the determination of aldehydes and ketones in ambient air.
High performance liquid chromatography(HPLC)Ambient airAcarbonyl compounds2,4-dinitrophenylhydrazine