In this work,an automated microfluidic chip that uses negative pressure to sample and analyze solutions with high temporal resolution was developed.The chip has a T-shaped channel for mixing the sample with a fluorescent indicator,a flow-focusing channel for generat-ing droplets in oil,and a long storage channel for incubating and detecting the droplets.By monitoring the fluorescence intensity of the droplets,the device could detect changes in solution accurately over time.The chip can generate droplets at frequencies of up to 42 Hz with a mixing ratio of 1∶1 and a temporal resolution of 3-6 s.It had excellent linearity in detecting fluorescein solution in the concentration range 1-5 μM.This droplet microfluidic chip provides several advantages over traditional methods,including high temporal resolution,stable droplet generation,and faster flow rates.This approach could be applied to monitoring calcium ions with a dynamic range from 102 to 107 nM and a detection limit of 10 nM.

Microfluidic chipDroplet samplingFluorescence detectionCalcium ion dynamicsTemporal resolution

Cong Ma、Zehang Gao、Jianlong Zhao、Shilun Feng

展开 >

School of Information Science and Technology,ShanghaiTech University,Shanghai 201210,China

Shanghai Institute of Microsystem and Information Technology,Chinese Academy of Sciences,Shanghai 200050,China

Center of Materials Science and Optoelectronics Engineering,University of Chinese Academy of Sciences,Beijing 100049,China

Department of Clinical Laboratory,Third Affiliated Hospital of Guangzhou Medical University,Guangdong 510150,China

展开 >

equipment research and development projects of the Chinese Academy of SciencesState Key Lab of Transducer Technology at Shanghai Institute of Microsystem and Information Technology,Chinese Academy of Sc