首页|国产SPECT/CT系统的定量性能研究

国产SPECT/CT系统的定量性能研究

扫码查看
通过与国际先进的单光子发射计算机断层扫描/电子计算机断层扫描(single-photon emission computed tomography/computed tomography,SPECT/CT)成像系统 GE Discovery NM/CT 670 对比,评估 了国产 SPECT/CT 成像系统 Insight NM/CTPro定量精度,以实现定量成像.该国产成像系统由北京永新医疗设备有限公司研制.该文通过精确定量标定Insight NM/CT Pro,结合中国医药行业推荐标准YY/T 1546-2017《用于SPECT成像CT衰减校正的试验方法》和美国电气制造商协会(National Electrical Manufactures Association,NEMA)标准 NEMA NU 1-2018《Gamma 相机的性能测量》的相关要求,采用低能高分(low energy high resolution,LEHR)准直器评估放射性核素锝-99 m(99mTc)成像的主要性能指标,并与GE Discovery NM/CT 670进行对比.首先,使用衰减精度模型测试空气、水和骨3种不同插件的衰减误差;其次,使用 NEMA NU 1-2018体模测试2个成像系统的定量精度.结果表明:Insight NM/CT Pro获得的重建图像中空气、水和骨的衰减误差依次为7.84%、8.38%和4.66%;直径为13、17、22和28 mm的"热区"小球的定量恢复系数分别为38.22%、50.98%、66.55%和71.32%,背景浓度偏差为7.95%.国产Insight NM/CT Pro成像系统获得的重建图像质量和定量精度与GE Discovery NM/CT 670成像系统效果相当,证实了该系统定量可靠.
Study on quantification performance evaluation of a domestic SPECT/CT system
[Objective]Single-photon emission computed tomography(SPECT)is still considered a nonquantitative imaging modality because it cannot perform attenuation correction.To acquire a quantitative SPECT image,Beijing Novel Medical Equipment Co.,Ltd.developed a domestic SPECT/CT system,Insight NM/CT Pro.It was equipped with dual digital detectors and enabled various scanning geometries.In addition to better spatial resolution,the system can correct attenuation effects in SPECT and realize quantitative reconstruction using CT.In this study,we evaluated the imaging accuracy of this domestic SPECT/CT system and compared it with advanced abroad systems.[Methods]To obtain quantitatively correct results,physical effects,including collimator blurring,object attenuation,scatter,and radionuclide decay,were modeled and corrected in the reconstruction algorithm.Moreover,a large cylindrical phantom was employed to obtain the calibration factor for the system so that we could convert the reconstruction result to a quantitative image in terms of Bq/mL.Then,the performance of CT-based attenuation correction was tested according to testing method for SPECT imaging based on CT-attenuation correction(YY/T 1546-2017).A cylinder phantom with three cylindrical inserts corresponding to air,nonradioactive water,and bone was filled with a radioactive solution for imaging.The biases inside the air,water,and bone regions of the reconstructed image were calculated to evaluate the performance of CT-based attenuation correction.In addition,the quantitative accuracy of the equipment was tested using the Notional Electrical Manufactures Association(NEMA)torso phantom according to performance measurements of Gamma cameras(NEMA NU 1-2018).Six fillable spheres with different diameters were set as targets for recovery evaluation,and the target-to-background concentration ratio was approximately 8:1.A large volume of interest(VOI)was placed in the background region to calculate the quantitative bias of the reconstruction.VOIs with the same size of six spheres were drawn based on the registered CT to evaluate recovery coefficients of different sizes.Radionuclide technetium-99m and a low-energy,high-resolution collimator with Insight NM/CT Pro were used for all these tests,and the evaluation results were compared with those of the GE Discovery NM/CT 670.[Results]For attenuation accuracy testing,the biases inside the air,water,and bone regions of Insight NM/CT Pro were 7.84%,8.38%,and 4.66%,respectively,whereas the corresponding biases in GE Discovery NM/CT 670 were 16.64%,18.01%,and 11.02%,respectively.For quantitative accuracy testing,in GE discovery NM/CT 670,the quantitative recovery coefficients of the hot spheres with diameters of 13,17,22,and 28 mm were 31.04%,51.36%,59.91%,and 66.39%,respectively,and the background concentration bias was 10.84%.Insight NM/CT Pro achieved higher recovery coefficients and lower bias.The recovery coefficients were 38.22%,50.98%,66.55%,and 71.32%for 13,17,22,and 28 mm hot rods,respectively,and the bias in the background region was 7.95%.[Conclusions]Phantom studies have demonstrated that the domestic Insight NM/CT Pro imaging system can obtain smaller biases after CT attenuation correction and achieve quantitative images with high accuracy.The reliability of the Insight NM/CT Pro system for quantitative imaging is validated in this study,and its performance is comparable to that of the advanced GE Discovery NM/CT 670 imaging system.

single-photon emission computed tomography/computed tomographyattenuation correctionphantom evaluationquantitative accuracy

江年铭、刘帆、程李、高丽蕾、刘辉、刘亚强

展开 >

清华大学工程物理系,北京 100084

北京永新医疗设备有限公司,北京 102206

SPECT/CT 衰减校正 模体评估 定量精度

国家重点研发计划专项

2022YFC2403800

2024

清华大学学报(自然科学版)
清华大学

清华大学学报(自然科学版)

CSTPCD北大核心
影响因子:0.586
ISSN:1000-0054
年,卷(期):2024.64(8)