The Influence of Hydrochloric Acid and Nitric Acid Solution System on ICP-MS Analysis of Six Precious Metal Elements
The paper used inductively coupled plasma mass spectrometry to determine six precious metal elements,Ru,Rh,Pd,Ir,Pt,and Au,in hydrochloric acid and nitric acid solution systems.The effects of acid type,acidity,and storage time on the linear correlation coefficient,linear error,memory effect,and spiked recovery rate of each element were studied.The results indicated that the 1%hydrochloric acid and nitric acid solution system was suitable for ICP-MS analysis of Ru,Rh,Pd,Ir,Pt,while the 3%to 4%hydrochloric acid solution system was suitable for ICP-MS analysis of Au elements.
hydrochloric acidnitric acidprecious metal elementsICP-MS