首页|智能一体化COD消解仪测定高氯废水COD的探讨

智能一体化COD消解仪测定高氯废水COD的探讨

Discussion on the Determination of COD in High Chlorine Wastewater by Intelligent Integrated COD Digestion Instrument

扫码查看
本文采用智能一体化COD消解仪,通过测定方法的检出限、测定下限、精密度和准确度,验证了该仪器在高氯废水COD测定中的可行性.实验结果表明:该仪器测定高氯废水COD检出限小于10mg/L,精密度相对标准偏差范围为1.3%~6.7%,准确度相对误差在-4.1%~2.5%,显示出较高的精密度和准确度,均符合标准(HJ/T 70-2001)的要求,并且操作简便.
The present study utilizes an intelligent integrated COD digestion instrument to assess the feasibility of measuring COD in high-chloride wastewater.This assessment is based on the detection limit,lower limit of quantification,precision,and accuracy of the measurement method.The experimental results indicate that the detection limit is less than 10 mg/L,with a relative standard deviation for precision ranging from 1.3%to 6.7%,and a relative error for accuracy between-4.1%and 2.5%.Both precision and accuracy demonstrate relatively high levels.The measurements obtained using the intelligent integrated COD digestion instrument meet the standards set forth in HJ/T70-2001 regarding precision and accuracy for determining COD in high-chloride wastewater.Furthermore,this instrument offers ease of operation.

intelligent integrationhigh chlorine wastewatermeasurement method

莫勇媚、陈广辉、刘文亮

展开 >

广东省东莞生态环境监测站,广东 东莞 523000

东莞市生态环境监控中心,广东 东莞 523000

智能一体化COD消解仪 高氯废水 测定方法

2024

山西化工
山西省煤化工发展促进中心 山西省化工学会

山西化工

影响因子:0.293
ISSN:1004-7050
年,卷(期):2024.44(12)