首页|智能一体化COD消解仪测定高氯废水COD的探讨

智能一体化COD消解仪测定高氯废水COD的探讨

扫码查看
本文采用智能一体化COD消解仪,通过测定方法的检出限、测定下限、精密度和准确度,验证了该仪器在高氯废水COD测定中的可行性.实验结果表明:该仪器测定高氯废水COD检出限小于10mg/L,精密度相对标准偏差范围为1.3%~6.7%,准确度相对误差在-4.1%~2.5%,显示出较高的精密度和准确度,均符合标准(HJ/T 70-2001)的要求,并且操作简便.
Discussion on the Determination of COD in High Chlorine Wastewater by Intelligent Integrated COD Digestion Instrument
The present study utilizes an intelligent integrated COD digestion instrument to assess the feasibility of measuring COD in high-chloride wastewater.This assessment is based on the detection limit,lower limit of quantification,precision,and accuracy of the measurement method.The experimental results indicate that the detection limit is less than 10 mg/L,with a relative standard deviation for precision ranging from 1.3%to 6.7%,and a relative error for accuracy between-4.1%and 2.5%.Both precision and accuracy demonstrate relatively high levels.The measurements obtained using the intelligent integrated COD digestion instrument meet the standards set forth in HJ/T70-2001 regarding precision and accuracy for determining COD in high-chloride wastewater.Furthermore,this instrument offers ease of operation.

intelligent integrationhigh chlorine wastewatermeasurement method

莫勇媚、陈广辉、刘文亮

展开 >

广东省东莞生态环境监测站,广东 东莞 523000

东莞市生态环境监控中心,广东 东莞 523000

智能一体化COD消解仪 高氯废水 测定方法

2024

山西化工
山西省煤化工发展促进中心 山西省化工学会

山西化工

影响因子:0.293
ISSN:1004-7050
年,卷(期):2024.44(12)