Objective To evaluate the precision and accuracy level of four Dimai DM79x blood cell analyzers in the most suitable working state of our department's daily indoor quality control by comparing the indoor quality control data,so as to improve the accuracy and consistency of test results.Methods In February 2023(28 days),the indoor quality control data of four laboratory instruments using original low,medium and high fixed value quality control products and daily testing of a random outpatient fresh patient sample were collected.The items included:White blood cells(WBC),red blood cells(RBC),hemoglobin(HGB),hematocrit(HCT),platelets(PLT),mean red blood cell volume(MCV),mean red blood cell hemoglobin volume(MCH),mean red blood cell hemoglobin concentration(MCHC)calculated mean(X),standard deviation(S)and coefficient of variation(CV),Coefficient of variation index(CVI)and standard deviation index(SDI)were used as precision and accuracy evaluation indexes respectively.Result The statistical analysis showed that the blood cell analysis of the DM79x instrument included 8 parameters,including WBC,RBC,HGB,PLT,HCT,MCV,MCH and MCHC.The coefficient of variation index of CVI of each instrument was different,and the indoor quality control result of SDI was<2.0.The measured values of CVI and SDI of fresh sample comparison data are all within the required range.Conclusion Four Dimai DM79x automatic blood cell analyzers in our department have good precision and high accuracy by comparing the indoor quality control data between laboratories.The performance of the instrument can meet the work needs of clinical laboratories.Moreover,the accuracy and consistency of the results in different instruments can be ensured by comparing the fresh samples of patients.