冷消解测定血清碘方法与国标法的比较分析

Comparative analysis of serum iodine determination by cold digestion method and national standard method

扫码查看
目的 评价冷消解测定血清碘方法在临床应用中的优势和价值.方法 应用长沙塞克陆德的DAT-500 型碘元素综合检测仪和冷消解血清碘检测试剂(专利:CN113340888A用于血清定量检测的试剂、试剂盒及检测方法)与国标《血清中碘的测定砷铈催化分光光度法》方法学适用性实验对比.结果 0~300 µg/L碘含量区间,5 批次标准曲线相关系数绝对值超 0.9995,检出限 2.1 µg/L;9 份不同浓度血清样加标回收率 94.6%~108.8%;3 种浓度血清样各测 6 批次,RSD为 3.41%~4.01%,样品加标回收率 94.6%~108.8%;20 例血清样本测定值与国标法差异无统计学意义(t=-1.660,P=0.113>0.05).结论 冷消解法检测血清碘精度与准度高,操作简便耗时短、温控要求低,试剂常温易存运,应用前景佳.
Objective To evaluate the advantage clinical value of cold digestion method and national standard method in the determination of serum iodine.Methods The DAT-500 iodine detector of Changsha Seclude and the cold digested serum iodine detection reagent(patent:CN113340888A reagent kit and detection method for serum quantitative detection)were used to compare with the national standard Determination of Iodine in Serum by Catalytic Spectrophotometry.Results From 0 to 300 μg/L iodine content range,the absolute value of correlation coefficient of standard curve of 5 batches exceeded 0.9995,the detection limit was 2.1 μg/L;the recovery rate of 9 serum samples with different concentrations was 94.6%~108.8%;6 batches of 3 serum samples with different concentrations were tested,the RSD was 3.41%~4.01%,and the recovery rate was 94.6%~108.8%;there was no statistical significance between 20 serum samples and national standard method(t=-1.660,P=0.113>0.05).Conclusion Cold digestion method has the advantages of high accuracy and accuracy,simple operation,short time consuming,low demand for temperature control,easy storage and transportation at room temperature.

cold digestion serum iodine detection reagentserum iodine automatic detection instrumentserum iodine detection

陈贵莲、丁黎、张颖、李春春、王晓

展开 >

美年大健康产业(集团)有限公司长沙公司,长沙 410000

长沙塞克陆德医疗科技有限公司,长沙 410000

冷消解血清碘检测试剂 血清碘自动检测仪 血清碘检测

2024

实验室检验
中国检验检测学会

实验室检验

ISSN:2097-261X
年,卷(期):2024.2(12)