Uncertainty Analysis and Experimental Study of Micro-angle Measuring Device
In order to improve the measuring accuracy of the micro-angle measuring device based on Fabry-Perot(F-P)etalon,two methods of guiding by the expression of measurement(GUM)and Monte Carlo method(MCM)were used to study the measuring uncertainty,respectively.GUM uncertainty evaluation model and MCM mathematical model were established.The initial position of the mirror was adjusted by the angle difference measurement method,and 150"measurement repeatability experiment was completed to obtain the measurement uncertainty of different evaluation methods.The evaluation results show that the uncertainty evaluation result of GUM is 0.037 5"and MCM is 0.040 0"within the 150"measurement range.Within the given tolerance range of 0.000 5",GUM is not verified by MCM,so MCM should be used to evaluate the measurement uncertainty in current and similar application scenarios.The normal probability plot method is used to test the normal distribution of the measured angle.It is found that the inconsistent results of the two evaluation methods are due to the deviation of the measured distribution function from the normal distribution.
micro-angle measurementmeasurement uncertaintyguiding by the expression of uncertainty in measurement(GUM)Monte Carlo method(MCM)Fabry-Perot(F-P)etalon