中国组织工程研究2025,Vol.29Issue(10) :2171-2177.DOI:10.12307/2025.254

口腔种植修复中钛基底氧化锆基台的制作和力学性能

Manufacture and mechanical property on zirconia abutments with a titanium base in dental implant restoration

王欢 陆婧 李英 孟茂花 舒佳玉 罗云彩 李文杰 董强
中国组织工程研究2025,Vol.29Issue(10) :2171-2177.DOI:10.12307/2025.254

口腔种植修复中钛基底氧化锆基台的制作和力学性能

Manufacture and mechanical property on zirconia abutments with a titanium base in dental implant restoration

王欢 1陆婧 1李英 1孟茂花 1舒佳玉 1罗云彩 1李文杰 1董强1
扫码查看

作者信息

  • 1. 贵州医科大学口腔医学院/贵州医科大学附属口腔医院,贵州省贵阳市 550004
  • 折叠

摘要

背景:随着计算机辅助设计和计算机辅助制造技术的发展,钛基底氧化锆基台因具备良好的应用优势被广泛运用于临床,但仍存在一些问题,尚缺乏共识性的设计标准.目的:综述钛基底氧化锆基台的制作方法及基台连接、穿龈设计、基台角度和粘接对钛基底氧化锆基台各类力学性能的影响.方法:检索PubMed数据库和中国知网2010-2023年发表的相关文献,检索词为"zirconia abutment,titanium base;氧化锆基台,钛基底",部分经典文献延长检索时间限制,通过纳入和排除标准获取相关文献,对57篇符合标准的文献进行综述.结果与结论:建议临床医生尽量选择具有抗旋结构的基台连接或是具有莫氏锥度的非抗旋连接,当运用钛基底氧化锆基台时,种植体应放置在正确的轴向角度(植体轴向向唇侧倾斜角度不超过15°或略向腭侧倾斜);当植体轴向向唇侧倾斜角度≥30°时,应当适当降低咬合力,以降低种植体、基台及修复体机械并发症和生物并发症的风险;应尽可能选择穿龈高度较高的钛基底,并且其穿龈角度不应超过40°.Multilink Hybrid Abutment可以作为氧化锆基台与钛基底口外粘结的首选粘接剂.

Abstract

BACKGROUND:With the development of computer-aided design and computer-aided manufacturing technology,zirconia abutments with a titanium base are widely used in clinic due to its good application advantages,but there are still some problems and a lack of consensus design standards.OBJECTIVE:To review the fabrication methods of Ti-base zirconia abutment,and the effect of abutment connection,emergence design,abutment angle,and bonding on mechanical properties of Ti-base zirconia abutment.METHODS:Relevant literature published from 2010 to 2023 was searched in CNKI and PubMed databases with the search terms"zirconia abutment,titanium base"in Chinese and English,respectively.The search time limit was extended for some classical literature.The relevant literature was obtained through inclusion and exclusion criteria,and 57 eligible documents were included for review.RESULTS AND CONCLUSION:It is recommended that clinicians try to select antirotational titanium bases or rotational titanium bases with a Morse taper connection.Implants should be placed in the correct axial angulation of not more than 15° or with an inclination to the palatal side when using angled zirconia abutments.When a≥30° labial inclination is followed for implant placement,the bite force must be decreased effectively to reduce the risk of mechanical and biological complications of implants,abutments,and prostheses.Ti-base zirconia abutments with a higher gingival height should be selected,and its restorative angle should not exceed 40°.Multilink Hybrid Abutment could be the first choice for extraoral bonding of zirconia abutment to titanium bases.

关键词

氧化锆基台/Ti-base/口腔种植/力学性能/钛基底/个性化基台/计算机辅助设计和计算机辅助制造/数字化

Key words

zirconia abutment/Ti-base/dental implant/mechanical property/titanium base/custom implant abutment/computer-aided design and computer-aided manufacturing/digital

引用本文复制引用

出版年

2025
中国组织工程研究
中国康复医学会,《中国组织工程研究与临床康复》杂志社

中国组织工程研究

CSTPCD北大核心
影响因子:1.387
ISSN:2095-4344
参考文献量57
段落导航相关论文