首页|2023年广东省数字减影血管造影透视防护区的周围剂量当量率分析

2023年广东省数字减影血管造影透视防护区的周围剂量当量率分析

扫码查看
目的 了解2023年广东省部分数字减影血管造影(Digital Subtraction Angiography,DSA)透视防护区的周围剂量当量率结果,为介入职业人员的辐射安全提供建议。方法 收集两家第三方放射卫生技术服务机构2023年中对142台DSA透视防护区的周围剂量当量率结果检测数据,并进行统计分析。结果 第一术者位和第二术者位的防护合格率比较,差异无统计学意义(P>0。05)。第一术者位和第二术者位下肢周围剂量当量率比较,差异无统计学意义(Z=-1。381,P=0。167)。第一术者位和第二术者位足部、腹部、胸部和头部周围剂量当量率比较,差异有统计学意义(P<0。05)。其中,第一术者位的足部检测结果高于第二术者位(Z=-3。271,P=0。001),腹部、胸部、头部检测结果低于第二术者位(Z=9。477、10。004、9。471,P<0。001)。结论 WS76-2020标准下DSA透视防护区的周围剂量当量率合格率较高,第二术者位监测结果高于第一术者位,介入诊疗放射工作人员仍须正确穿戴个人防护用品,合理使用辅助防护设施,以保证职业健康。
Analysis of the ambient dose equivalent rate in test surface fluoroscopy protection area of digital subtraction angiography devices in Guangdong in 2023
Objective To understand the result of the ambient dose equivalent rate in test surface of Digital Subtraction Angiography(DSA)device fluoroscopy protection area in Guangdong in 2023,and to provide suggestions for radiation safety for interventional occupational personnel.Methods Collect and analyze the data of 142 DSA devices the ambient dose equivalent rate in test surface that from two Radiation Health Technology Service Institutions.Results The qualification rate of the first operator position and the second operator position in terms of protection was compared,and there was no significant difference in terms of statistics(P>0.05).The equivalent dose rate of the lower limbs in the first operator position and the second operator position was compared,and there was no significant difference in terms of statistics(Z=-1.381,P=0.167).The equivalent dose rate of the foot,abdomen,chest,and head in the first operator position and the second operator position was compared,and there was a significant difference in terms of statistics(P<0.05).Among them,the detection results of the foot in the first operator position were higher than those in the second operator position(Z=-3.271,P=0.001),and the detection results of the abdomen,chest,and head were lower than those in the second operator position(Z=9.477,10.004,9.471,P<0.001).Conclusions According to the methods of WS76-2020,the qualified rate of the ambient dose equivalent rate in test surface of DSA fluoroscopy protection area was high.In order to protect the occupational health of interventional radiation workers,they are required to wear lead aprons correctly and use auxiliary protective facilities reasonably.

Digital Subtraction AngiographyAmbient Dose Equivalent RateRadiological ProtectionOccupational Health

程希元、张德天、苏冠铭

展开 >

广州南医宜康医疗设备股份有限公司,广州 510515

广州南方医大医疗设备综合检测有限责任公司,广州 510515

广州南大康德信检测技术有限责任公司,广州 510515

数字减影血管造影 周围剂量当量率 放射防护 职业健康