首页|西门子Force双源CT故障维修案例分析与探讨

西门子Force双源CT故障维修案例分析与探讨

扫码查看
西门子Force双源CT属于高端CT,设备整体配置高、性能佳、应用范围广、稳定性好,在医院运行中发挥了巨大作用.由于设备使用量大,故障时有发生,且设备集成程度较高、整体维修难度大.本文详细分析了使用过程中发生的水冷系统、患者检查床、探测器电源3例故障的故障排除过程,并通过更换三通阀、维护升降电机轴承、焊接X10接头解决故障,节省了维修费用,并为同行维修此类故障提供参考.
Analysis and Discussion of Siemens Force Dual Source CT Fault Maintenance Case
Siemens Force dual source CT is a high-end CT,with high overall configuration,good performance,wide application range,and good stability,playing a huge role in the operation of hospital.Due to the high usage of equipment,failures have occurred from time to time,and the high level of equipment integration makes overall maintenance difficult.This paper analyzed in detail the troubleshooting process of three cases of water cooling system,patient inspection bed and detector power supply faults occurred during use,and solved the faults by replacing the three-way valve,maintaining the lifting motor bearing and welding X10 joint,which can save maintenance costs and provide reference for peers to repair such faults.

dual source CTwater cooling systempatient tabledetector power supply

孙小磊、肖吴进、汪缨、张晖

展开 >

南京医科大学第一附属医院 临床医学工程处,江苏 南京 210029

双源CT 水冷系统 检查床 探测器电源

2024

中国医疗设备
中国整形美容协会

中国医疗设备

CSTPCD
影响因子:0.825
ISSN:1674-1633
年,卷(期):2024.39(8)