Analysis and Discussion of Siemens Force Dual Source CT Fault Maintenance Case
Siemens Force dual source CT is a high-end CT,with high overall configuration,good performance,wide application range,and good stability,playing a huge role in the operation of hospital.Due to the high usage of equipment,failures have occurred from time to time,and the high level of equipment integration makes overall maintenance difficult.This paper analyzed in detail the troubleshooting process of three cases of water cooling system,patient inspection bed and detector power supply faults occurred during use,and solved the faults by replacing the three-way valve,maintaining the lifting motor bearing and welding X10 joint,which can save maintenance costs and provide reference for peers to repair such faults.
dual source CTwater cooling systempatient tabledetector power supply