总结了费森尤斯4008系列血液透析机2年的故障维修记录,分析了流量报警、漏水和漏气、堵塞、机械性损坏等各类型故障产生的主要原因,基于DMAIC方法提出了优化维修、保养流程的建议,提高了保养覆盖率和设备稳定性,为透析临床技师合理分配维修保养时间、及早发现故障隐患提供了参考.
Application of DMAIC method in maintenance of hemodialysis machine
Two-year maintenance records of Fresenius 4008 series hemodialysis machines were summarized.The main causes were analyzed for types of faults including flow alarm,water and air leakage,clogging,mechanical damage and etc.Some suggestions based on DMAIC method were proposed to optimize the maintenance precedure and enhance maintenance coverage and machine stability.References were provided for hemodialysis technician to allocate maintenance time effectively and find hidden fault at an early stage.[Chinese Medical Equipment Journal,2025,46(1):78-82]

hemodialysis machineDMAIC methodtroubleshootingmaintenance