Determination of Chlorine Content in Dimethyl Dichlorosilane Hydrolysate by Combustion Ion Chromatography
A combustion ion chromatography method was established to determine the chlorine content in dimethyl hydrolysates.The detection limit of this method is 0.06mg/kg,the recovery rate is 80.0%~82.6%,and the relative standard deviation(RSD)is less than 5%.It can meet the determination of trace chlorine content in hydrolysates without the need for pre-treatment,and the sample detection is simple and conven-ient.It can be used for routine analysis of chlorine content in hydrolysates.
Combustion Ion ChromatographyDimethyldichlorosilane HydrolysateChlorine Content