中国保健营养(下旬刊)2014,Issue(5) :2969-2970,2971.DOI:10.3969/j.issn.1004-7484(x).2014.05.707

UF-100尿液有形成分分析仪酵母菌参数影响因素的探讨

Explored the Effecting of yeast Determined by UF-100 Automated Urinalysis Analyzed

牛小斌 李永伟
中国保健营养(下旬刊)2014,Issue(5) :2969-2970,2971.DOI:10.3969/j.issn.1004-7484(x).2014.05.707

UF-100尿液有形成分分析仪酵母菌参数影响因素的探讨

Explored the Effecting of yeast Determined by UF-100 Automated Urinalysis Analyzed

牛小斌 1李永伟1
扫码查看

作者信息

  • 1. 河南省中医院,河南 郑州 450002
  • 折叠

摘要

目的:探讨UF-100尿液有形成分分析仪酵母菌参数的准确性及其影响因素。方法对1160份尿液标本同时用UF-100尿液有形成分分析仪和离心镜检法进行酵母菌检测。结果 UF-100尿液有形成分分析仪对酵母菌检测的准确率为94.5%,阳性预测值为29.6%,结晶、红细胞、细菌、精子细胞等可造成酵母菌假阳性,高浓度的红细胞也可使酵母菌出现假阴性。结论 UF-100尿液有形成分分析仪检测酵母菌结果不能作为确定性报告,应通过镜检复审,以免发生误报。

Abstract

Objective To explore the accuracy and influencing factors of yeast determined by UF -100 automated urinalysis analyzer.Methods Urine specimens from 1160 patients were examined for yeast with UF -100 automated urinalysis analyzer and microscopic examination .Results The data obtained showed that the accura-cy of UF-100 automated urinalysis analyzer in yeast was 94.5%,positive predictive value was 29.6%;crystal,erythrocyte,bacteria and sperm could affect the accuracy of determination.Conclusion There are more false positive and low positive predictive value in yeast determined by UF -100 automated urinalysis analyzer,the combina-tion of microscopic examination of urine sediment can increase the accuracy of detection in yeast .

关键词

UF-100尿液有形成分分析仪/尿沉渣/显微镜/酵母菌

Key words

UF-100 automated urinalysis analyzer/Urine sediment/Microscope/Yeast

引用本文复制引用

出版年

2014
中国保健营养(下旬刊)
全国卫生产业企业管理协会

中国保健营养(下旬刊)

ISSN:1004-7484
被引量1
参考文献量4