针对医用电气设备的漏电流和患者辅助电流试验,本文以电气安全标准GB 9706.1-2020为基础,简要阐述了医用电气设备的产生原因并分析漏电流的危害,浅析医用电气设备漏电流的分类、测试网络的要求、单一故障状态,并从理论和实际操作上进行分析,介绍试验过程中需要注意的事项以及符合性判定,为有源医疗器械制造商和第三方检验检测机构的研发、测试人员正确理解该试验提供一定参考.
Discussion on Leakage Current Test of Medical Electrical Equipment
For the leakage current and patient auxiliary current testing of medical electrical equipment,the paper is based on the standard for electrical safety,GB 9706.1-2020,briefly expounds the causes of medical electrical equipment,and analyzes the hazards of leakage current.It analyzes the classification of leakage current in medical electrical equipment,test network requirements and single fault conditions.From both theoretical and practical perspectives,the paper introduces the precautions and compliance assessments required during the test process,and provides certain references for the R&D/testing personnel from active medical device manufacturers and third-party inspection and testing institutions to properly understand the test.

leakage currentpatient auxiliary currentsingle fault conditionnon-frequency weightingmedical electrical equipment

朱洪亮

展开 >

上海国研医疗器械检测中心有限公司

漏电流 患者辅助电流 单一故障状态 非频率加权 医用电气设备

2025

中国标准化
中国标准化协会 中国标准化研究院

中国标准化

影响因子:0.241
ISSN:1002-5944
年,卷(期):2025.(1)