Study on Determination of Sulfur Content in Diatomite by High Frequency Infrared Carbon-Sulfur Analyzer
The sulfur content in diatomite was measured by high frequency infrared carbon-sulfur analyzer(CS-8800C).When the sample size was 0.040~0.050 g,0.2g pure iron flux and 1.5 g pure tungsten flux were added,the detection limit is 0.0003%,the precision is 1.77%~4.52%,the accuracy is good,the analysis method is quick and simple,the direct weighing sample is rapid,the analysis result can satisfy the geological criterion.
high frequency infrared carbon-sulfur analyzerdiatomitesulfur