研究了直读式测钙仪水泥剂量的校准方法,阐明了直读式测钙仪原理,介绍了校准用标准溶液的配制,给出仪器示值误差和重复性的测定步骤,列出了示值误差测量不确定度评定方法.

Abstract

This paper studies the calibration method of cement dosage for direct reading calcium meter,elucidates the principle of direct reading calcium meter,introduces the preparation of standard solutions for calibration,provides the measurement steps for instrument indication error and repeatability,and lists the methods for evaluating the measurement uncertainty of indication error.

关键词

直读式测钙仪/水泥剂量/示值误差/重复性/不确定度

Key words

direct reading calcium meter/cement dosage/indication error/repeatability/uncertainty