中国激光2024,Vol.51Issue(5) :229-234.DOI:10.3788/CJL231426

差分式共振光声光谱技术研究

Study on Differential Resonance Photoacoustic Spectroscopy

徐靖 黄琪 李劲松
中国激光2024,Vol.51Issue(5) :229-234.DOI:10.3788/CJL231426

差分式共振光声光谱技术研究

Study on Differential Resonance Photoacoustic Spectroscopy

徐靖 1黄琪 2李劲松2
扫码查看

作者信息

  • 1. 航宇救生装备有限公司,湖北襄阳 441003
  • 2. 安徽大学物理与光电工程学院,安徽合肥 230601
  • 折叠

摘要

针对光声光谱系统本底噪声限制问题,以圆柱形声共振器作为研究对象,开展了双共振腔光声池设计,并将共振光声光谱和差分探测原理相结合,建立了一套基于近红外半导体激光器的高灵敏度激光光谱气体探测系统.以空气中的水汽(H2O)分子作为检测对象,结合高灵敏度波长调制二次谐波探测方法,对建立的差分式共振光声光谱系统进行了理论分析和实验评估.艾伦方差分析结果显示,系统可实现几个10-6量级水汽分子浓度的高灵敏度检测.相较于传统单通道光声探测模式,所提出的差分探测式共振光声光谱探测技术可有效提升系统的稳定性和检测灵敏度,最佳信号平均时间可提高2倍.

Abstract

Objective Photoacoustic spectroscopy(PAS)is a powerful and non-destructive optical analysis technique that can be used to quantitatively analyze the composition of gases,liquids,and solids.With the continuous innovation of modern laser techniques,various new laser light sources have emerged that play an important role in promoting the development of photoacoustic spectroscopy based on laser light sources.Moreover,novel signal processing algorithms and detection strategies have been reported.In theory,PAS is essentially established with wavelength independence,high resolution,and high sensitivity.These unique characteristics make them widely used in environmental science,solid-state physics,industrial process control,biomedicine,and other fields.However,the thermal noise caused by the absorption of the incident laser by the windows or inner wall of the photoacoustic(PA)cell(particularly when the high-power laser source is used as the signal excitation light source),and the electrical noise of electronic devices(such as acoustic signal detectors)are still key technical issues that limit the detection sensitivity of PA-spectroscopy-based gas sensors.To resolve the background noise problem in these technical issues,a differential resonance photoacoustic gas detection method that fully utilizes the phase-dependent characteristics of the PA resonance cavity is proposed.Methods Considering the problem of noise limitation in PAS system sensitivity,resonance enhancement detection strategies are usually adopted to achieve effective suppression of the background noise of the PA system.Typically,cylindrical resonant PA cells,Helmholtz resonators,spherical resonators,and quartz tuning forks are used.In the field of signal processing algorithms,differential detection is an effective method for eliminating background noise interference,improving signal quality,and improving the spectral signal-to-noise ratio(SNR),and has good application value in various signal processing.Therefore,in this study,a high-sensitivity PA gas detection technique is developed by combining resonance PAS characteristics based on the differential detection principle.To demonstrate the proposed technique,a cylindrical PA cell with double resonant cavities is designed,and relevant theoretical and experimental studies are conducted for sensitive sensing gas detection.A differential-resonance PAS gas sensor system is integrated by using a near-infrared diode laser near 1391.6 nm and a double PA cell.To further improve the detection sensitivity,a wavelength modulation spectroscopy second-harmonic(WMS-2F)detection method is employed.Moreover,the Allan variance analysis algorithm is used to evaluate the system sensitivity and stability.Results and Discussions To evaluate the gas-sensing technique,ambient water vapor(H2O)is analyzed.The potential crosstalk effect between the double-resonance cavities is investigated using theoretical simulations(Fig.1)and experimentally confirmed.Differential detection is applied for measuring background noise and H2O PA spectral signals.The calculated results indicate that the standard deviations of the background noise can be improved by approximately 1.9 times(Fig.4)by utilizing the phase-dependent characteristics of the two resonance cavities(Fig.5),and the PA spectral signal amplitude can also be significantly enhanced(Fig.7).Moreover,a detection limit of~3.0× 10-6 is obtained for ambient H2O concentration measurements under the optimal averaging time of 115 s without using differential detection(Fig.8).After using the differential algorithm,the system stability is further improved,the optimal stability time is increased to more than 200 s,and the corresponding detection sensitivity is improved to 2.0×10-6(Fig.8).Conclusions This study proposes a high-sensitivity gas detection technique based on resonant PAS with differential detection principle.Allan variance analysis indicates that high-sensitivity detection of several 10-6 level H2O concentrations can be achieved using a low-power near-infrared(NIR)diode laser.Compared to the traditional single-channel PA detection mode,the results prove that the proposed differential resonant PAS detection technique can effectively improve the system stability and detection sensitivity,and the optimal signal average time can be doubled.

关键词

光谱学/光声光谱/差分探测/共振效应/气体检测

Key words

spectroscopy/photoacoustic spectroscopy/differential detection/resonant effect/gas detection

引用本文复制引用

基金项目

国家自然科学基金(41875158)

国家自然科学基金(61675005)

出版年

2024
中国激光
中国光学学会 中科院上海光机所

中国激光

CSTPCDCSCD北大核心
影响因子:2.204
ISSN:0258-7025
参考文献量19
段落导航相关论文