日本(包括国产)的ECG-6511型、ECG-6151型两种型号的心电图机广泛应用于医疗卫生单位.基于此,本文主要对心电图机记录笔偏移、上下摆动、颤抖等部分故障进行详细分析,提出具体维修方法.相关人员应定期对心电图机进行维护和检查,确保其处于良好的工作状态,使用高质量的导联线和配件,减少因配件问题导致的故障.对操作人员进行培训,确保相关人员正确规范地使用心电图机.

Abstract

The ECG-6511 and ECG-6151 electrocardiographs from Japan(including domestic ones)are widely used in medical and health institutions.Based on this,this article mainly analyzes in detail the faults of the recording pen offset,up and down swinging,and trembling of the electrocardiogram machine,and proposes specific maintenance methods.Relevant personnel should regularly maintain and inspect the electrocardiogram machine to ensure that it is in good working condition.Use high-quality lead wires and accessories to reduce malfunctions caused by accessory issues.Train operators to ensure that they can use the electrocardiogram machine correctly and in a standardized manner.

关键词

心电图机/故障/维修方法

Key words

electrocardiograph/malfunction/maintenance method