中国临床医学影像杂志2024,Vol.35Issue(11) :814-818.DOI:10.12117/jccmi.2024.11.012

125I粒子植入术后腹部CT金属伪影抑制的iMAR最佳预设参数研究

The optimal preset of iMAR to reduce the metal artifact of 125I seeds in abdominal CT

李焕焕 孙文阁
中国临床医学影像杂志2024,Vol.35Issue(11) :814-818.DOI:10.12117/jccmi.2024.11.012

125I粒子植入术后腹部CT金属伪影抑制的iMAR最佳预设参数研究

The optimal preset of iMAR to reduce the metal artifact of 125I seeds in abdominal CT

李焕焕 1孙文阁1
扫码查看

作者信息

  • 1. 中国医科大学附属第一医院放射科,辽宁 沈阳 110001
  • 折叠

摘要

目的:研究降低125I粒子植入术后腹部CT去金属伪影迭代算法(iMAR)的最佳预设参数.方法:纳入2023年1月—2024年4月在我院行腹部125I粒子植入手术的患者17例.其腹部CT图像使用8种目前设备提供的iMAR预设进行后处理.在每组图像中,分别把感兴趣区放置在受125I粒子伪影影响的邻近组织区、125I粒子周围最明显的低密度伪影区、高密度伪影区和对侧腰大肌区.每组图像的伪影指数(AI)、含伪影组织ROI内的标准差值(SD)及主观评分值与没有使用iMAR(iMARNO)图像的测量值进行组间统计分析.结果:在8种预设参数中,颅内金属圈(NC)预设参数组的图像的AI值、SD值最低.在主观评分中NC预设参数组图像也得到了最高的主观评分值.结论:在iMAR的8种预设参数中,NC预设参数对减少腹部125I粒子植入术后CT成像金属伪影的效果最为显著.

Abstract

Objective:To investigate the optimal preset parameters of iterative metal artifact reduction (iMAR) for reducing metal artifacts in abdominal CT after 125I seeds implantation.Methods:Seventeen patients who underwent abdominal 125I seeds implantation surgery in our hospital from January 2023 to April 2024 were included.The abdominal CT images were post processed using eight currently available iMAR presets.In each set of images,the region of interest is placed in the adjacent tissue area affected by the 125I seeds,the low-density artifact area around the 125I seeds,the high-density artifact area,and the contralateral vertebral muscle.The artifact index (AI),standard deviation (SD) within the ROI and subjective scoring values of each group of images were compared with images without iMAR (iMARNO).Results:Among the eight presets,the AI and SD values of the images in the neuro-coils (NC) preset parameter group were the lowest.In subjective evaluation,the NC preset group image also obtained the highest subjective evaluation value.Conclusion:Among the eight presets for iMAR,the NC preset has the most significant effect on reducing CT imaging artifacts after abdominal 125I seeds implantation.

关键词

碘苷/腹部/体层摄影术,螺旋计算机

Key words

Idoxuridine/Abdomen/Tomography,Spiral Compued

引用本文复制引用

出版年

2024
中国临床医学影像杂志
中国医学影像技术研究会,中国医科大学

中国临床医学影像杂志

CSTPCDCSCD北大核心
影响因子:1.204
ISSN:1008-1062
段落导航相关论文