首页|Imaging plate scanners calibration and the attenuation behavior of imaging plate signals
Imaging plate scanners calibration and the attenuation behavior of imaging plate signals
扫码查看
点击上方二维码区域,可以放大扫码查看
原文链接
NETL
NSTL
万方数据
Based on previously reported work,we propose a new method for calibrating image plate(IP)scanners,offering greater flexibility and convenience,which can be extended to the calibration tasks of various scanner models.This method was applied to calibrate the sensitivity of a GE Typhoon FLA 7000 scanner.Additionally,we performed a calibration of the spontaneous signal attenuation behavior for BAS-MS,BAS-SR,and BAS-TR type IPs under the 20±1 ℃ environmental conditions,and observed significant signal carrier diffusion behavior in BAS-MS IP.The calibration results lay a foundation for further research on the interaction between ultra-short,ultra-intense lasers and matter.