Imaging plate scanners calibration and the attenuation behavior of imaging plate signals

扫码查看
Based on previously reported work,we propose a new method for calibrating image plate(IP)scanners,offering greater flexibility and convenience,which can be extended to the calibration tasks of various scanner models.This method was applied to calibrate the sensitivity of a GE Typhoon FLA 7000 scanner.Additionally,we performed a calibration of the spontaneous signal attenuation behavior for BAS-MS,BAS-SR,and BAS-TR type IPs under the 20±1 ℃ environmental conditions,and observed significant signal carrier diffusion behavior in BAS-MS IP.The calibration results lay a foundation for further research on the interaction between ultra-short,ultra-intense lasers and matter.

image platescannerultra-short ultra-intense lasers