生物安全柜高效过滤器检漏时上游气溶胶浓度的确定

扫码查看
目的 探讨生物安全柜高效过滤器完整性检测中的适宜上游气溶胶浓度.方法 依据YY 0569-20116.3.2《Ⅱ级生物安全柜》相关检测方法,在一台生物安全柜高效过滤器的上风侧施放10~91 μg/L的PAO气溶胶,在下游检测漏过率,研究漏过率与上游气溶胶浓度的相关性.结果 上游气溶胶浓度在10~50 μg/L时,漏过率与上游气溶胶浓度呈负相关,送风高效过滤器漏过率范围为0.000 1%~0.002 1%,排风高效过滤器漏过率范围为0.000 1%~0.005 3%.上游气溶胶浓度>50 μg/L时,最大漏过率不再变化.结论 生物安全柜高效过滤器检漏过程中上游气溶胶浓度在10~30 μg/L较为适宜.
Determination of upstream aerosol concentration in leak detection of high efficiency particulate air filter in biosafety cabinet
Objective To investigate the appropriate upstream aerosol concentration in the integrity test of the high efficiency particulate air(HEPA)filter of the biosafety cabinet.Methods According to the method of YY 0569-20116.3.2 of Class II biological safety cabinet,different concentration of PAO aerosol ranged from 11-91 μg/L were applied to the upwind side of the HEPA filter of a biological safety cabinet,and the leakage rates were detected downstream,and the correlation between leakage rate and upstream aerosol concentration was studied.Results When the upstream aerosol concentrations ranged from 11-50 μg/L,the leakage rates were negatively correlated with the upstream aerosol concentration,and the leakage rate of the supply air HEPA filter ranged from 0.000 1%-0.002 1%,and the leakage rate of the exhaust HEPA filter ranged from 0.000 1%-0.005 3%.At the upstream aerosol concentration>50 μg/L,the maximum leak rate no longer changed.Conclusion The upstream aerosol concentration of 10-30 μg/L is more appropriate in the leakage detection process of HEPA filter in biological safety cabinet.

HEPA filterintegrity detectionleakage rate

杨雪峰、吴磊、徐佩茹

展开 >

安徽省疾病预防控制中心,安徽合肥 230061

高效过滤器 完整性检测 漏过率

2024

中国消毒学杂志
军事医学科学院疾病预防控制所,中华预防医学会

中国消毒学杂志

CSTPCD
影响因子:0.754
ISSN:1001-7658
年,卷(期):2024.41(2)
  • 13