首页|三级生物安全实验室及关键防护设备的消毒效果验证

三级生物安全实验室及关键防护设备的消毒效果验证

扫码查看
目的 验证某疾病预防控制中心三级生物安全实验室及关键防护设备的汽化过氧化氢消毒效果.方法 采用模拟现场消毒试验方法,在该生物安全实验室核心间及生物安全柜、独立通风笼具、风口式排风高效过滤单元和污水消毒设备等处布放嗜热脂肪杆菌芽孢菌片,经过汽化过氧化氢消毒机常规消毒程序进行消毒,对消毒效果进行验证与评价.结果 该生物安全实验室核心间及关键防护设备所有菌片上细菌芽孢全部被杀灭,验证结果全部合格.实验室内脉动真空压力蒸汽灭菌器经化学指示卡、生物指示物和B-D测试3项验证全部合格.正压生物防护头罩经现场消毒试验,对自然菌平均杀灭对数值>1.0.结论 对该三级生物安全实验室及关键防护设备进行消毒效果验证,均已达到规定的指标要求.
Validation of disinfection effectiveness for a level 3 biosafety laboratory and the key protective equipment
Objective To verify the disinfection effect of vaporised hydrogen peroxide in a level 3 biosafety laboratory and the key protective equipment of a Centre for Disease Control and Prevention.Methods Simulated field disinfection test method was used to verify and evaluate the disinfection effect.The tablets of Bacillus stearothermophilus spores as indicators were placed in the biosafety laboratory core room,biosafety cabinet,individual ventilation cages,vent-type exhaust HEPA unit and sewage disinfection equipment,which were detected after the routine disinfection procedure of vaporized hydrogen peroxide disinfector.Results All tablets of Bacillus stearothermophilus spores in the core room of the biosafety laboratory and the key protective equipment on were completely killed,the results were all qualified.The pulsating vacuum pressure steam sterilizer in the laboratory was all qualified by chemical indicator card,biological indicator and B-D test verified.The disinfection field test of positive pressure biological protective hood showed that the average killing logarithm of natural bacteria was above 1.0.Conclusion The disinfection effect of the level 3 biosafety laboratory and the key protective equipment has been verified,and the requirements of the specified indexes have been reached.

level 3 biosafety laboratorykey protective equipmentvaporized hydrogen peroxide disinfectordisinfection efficacy verification

李金华、刘瑞佳、孙长荣、张明昊、蔡森鹏、平安琪、张成斌、徐国鑫

展开 >

环安生物技术服务有限公司,北京 100070

三级生物安全实验室 关键防护设备 汽化过氧化氢消毒机 消毒效果验证

2024

中国消毒学杂志
军事医学科学院疾病预防控制所,中华预防医学会

中国消毒学杂志

CSTPCD
影响因子:0.754
ISSN:1001-7658
年,卷(期):2024.41(12)