Analysis of the Role of Automatic Blood Group Analyzer and Microplate Method in Irregular Antibody Screening for Blood Donors
Objective:To study the effect of automatic blood type analyzer and microplate method in screening irregular antibodies for blood donors.Methods:Taking the blood samples of 60 blood donors from January 2022 to January 2023 as an example,all samples were detected by microplate method(microplate group)and automatic blood type analyzer method(automatic group),and the screening results of the two groups were compared,and the blood types of lipid blood samples,hemolytic samples and irregular antibody types were judged.Results:In the automatic group,there were 2 cases of anti-E,3 cases of anti-P1,1 case of anti-I,1 case of anti-Mur and 3 cases of cold autoantibodies,but no agglutination was detected by microplate method.The correct rate of blood group judgment of hemolytic samples in automatic group was higher than that in microplate group(P<0.05).The correct rate of judging the blood type of lipemia samples in automatic group was higher than that in microplate group(P<0.05).The detection rates of irregular antibodies,O cell agglutination,IgM and IgG antibodies in automatic group were higher than those in microplate group(P<0.05).Conclusion:In screening and identifying irregular antibodies for blood donors,accurate results can be obtained.
blood donorsirregular antibody screeningautomatic blood type analyzermicroplate method