首页|低温过氧乙酸灭菌器对腔镜器械灭菌质量的影响研究

低温过氧乙酸灭菌器对腔镜器械灭菌质量的影响研究

扫码查看
目的:探究低温过氧乙酸灭菌器对腔镜器械灭菌质量的影响.方法:选择2022年9月~2022年11月本院消毒供应中心接收的需要清洗消毒灭菌100件腔镜器械为研究对象,随机分为观察组和对照组各50件.两组腹腔镜器械经过清洗、干燥后,观察组采用低温过氧乙酸灭菌器进行灭菌处理,对照组应用低温过氧化氢灭菌器灭菌.通过物理、化学以及微生物培养评价器械灭菌效果,并应用便携式多气体检测仪监测灭菌器灭菌过程中的环境浓度,比较两组灭菌所需时间.结果:观察组腔镜器械灭菌平均时间明显短于对照组(P<0.05).两组腔镜器械物理、化学监测灭菌合格率比较差异无统计学意义(P>0.05).观察组内镜镜头、镜体微生物培养检测合格率明显高于对照组(P<0.05).两组灭菌后空气残留的过氧乙酸、过氧化氢浓度均低于标准.结论:应用低温过氧乙酸灭菌器对腔镜器械进行灭菌处理能够获得良好的灭菌效果,可缩短灭菌时间提高工作效率,且其安全性良好.
Study on the Effect of Low Temperature Peroxyacetic Acid Sterilizer on the Sterilization Quality of Endoscopic Instruments
Objective:To explore the impact of low-temperature peracetic acid sterilizers on the sterilization quality of endoscopic instruments.Methods:100 laparoscopic instruments requiring cleaning,disinfection and sterilization received from the disinfection supply center of our hospital from September 2022 to November 2022 were selected as the study objects,and randomly divided into observation group and control group with 50 pieces each.After cleaning,disinfection,drying and packaging,the observation group was sterilized with low temperature peracetic acid sterilizer,and the control group was sterilized with low temperature hydrogen peroxide sterilizer.The sterilization effect of the instrument was evaluated through physical,chemical and microbial culture,and the portable multi-gas detector was used to monitor the environmental concentration in the sterilization process of the sterilizer,and the sterilization time of the two groups was compared.Results:The average sterilization time of laparoscopic instruments in observation group was significantly shorter than the control group(P<0.05).There were no significant difference in the qualified rate of sterilization in physical and chemical between the two groups(P>0.05).The qualified rate of microbial culture detection of endoscopic lens and endoscopic instruments in the observation group were significantly higher than the control group(P<0.05).The concentrations of peracetic acid and hydrogen peroxide in the air after sterilization were lower than the standard.Conclusion:The application of low temperature peracetic acid sterilizer to sterilize endoscopic instruments can obtain good sterilization effect,shorten sterilization time and improve working efficiency,and its safety is good.

laparoscopic instrumentlow temperature peracetic acid sterilizersterilization effectdisinfection supply center

张云英、叶碧华

展开 >

宁德师范学院附属宁德市医院消毒供应中心(福建宁德 352100)

腔镜器械 低温过氧乙酸灭菌器 灭菌效果 消毒供应中心

2024

中国医疗器械信息
中国医疗器械行业协会

中国医疗器械信息

影响因子:0.375
ISSN:1006-6586
年,卷(期):2024.30(4)
  • 14