首页|心电监护仪质量控制检测效果评价

心电监护仪质量控制检测效果评价

扫码查看
目的:加强心电监护仪的安全风险管理.方法:从2022年1月1日~2022年12月31日选取本院在用的313台心电监护仪作为研究对象,预防性维护保养项目根据工程师维护保养检查情况判断是否合格,质量控制检测采用Prosim 8多参数病人模拟器以及Prosim Spo2血氧饱和度质量检测仪检测心电监护仪的外观功能、心率、呼吸率、无创血压、血氧饱和度、声光报警、报警限检查、静音检查等,并对心电监护仪进行检测效果评价.结果:监护仪的外观清洁保养合格率为84.34%,内部清洁保养合格率为82.43%,更换易损件合格率为83.71%,功能检测合格率为92.97%,安全检查合格率为95.21%.心电监护仪外观及输出参数的质量控制检测,心率合格率为91.37%,呼吸率合格率为89.14%,无创血压合格率为86.58%,血氧饱和度合格率为86.62%,外观功能合格率为87.22%,声光报警合格率为91.37%,报警限检查合格率为95.53%,静音检查合格率为95.21%.结论:根据心电监护仪的预防保养和质量控制检测效果,针对心电监护仪的预防性维护提出相应对策,提高心电监护仪输出参数的准确性,降低医疗风险.
Evaluation of Quality Control Detection Effect of ECG Monitor
Objective:Strengthen the safety risk management of ECG monitor.Methods:From January 1,2022 to December 31,2022,313 electrocardiogram monitors in use in our hospital were selected as research objects.The preventive maintenance project was judged to be qualified based on the maintenance and inspection results of engineers.Quality control testing was conducted using the Prosim 8 multi parameter patient simulator and the Prosim Spo2 blood oxygen saturation quality detection instrument to detect the appearance function,heart rate,respiratory rate,non-invasive blood pressure,blood oxygen saturation,sound and light alarms of the electrocardiogram monitors alarm limit check,mute check,etc.,and evaluate the detection effect of the electrocardiogram monitor.Results:The qualified rate of external cleaning and maintenance of the monitor is 84.34%,the qualified rate of internal cleaning and maintenance is 82.43%,the qualified rate of replacing vulnerable parts is 83.71%,the qualified rate of functional testing is 92.97%,and the qualified rate of safety inspection is 95.21%.The quality control testing of the appearance and output parameters of the electrocardiogram monitor showed a heart rate qualification rate of 91.37%,respiratory rate qualification rate of 89.14%,non-invasive blood pressure qualification rate of 86.58%,blood oxygen saturation qualification rate of 86.62%,appearance function qualification rate of 87.22%,sound and light alarm qualification rate of 91.37%,alarm limit inspection qualification rate of 95.53%,and mute inspection qualification rate of 95.21%.Conclusion:Based on the preventive maintenance and quality control testing results of electrocardiogram monitors,corresponding measures are proposed for the preventive maintenance of electrocardiogram monitors to improve the accuracy of output parameters and reduce medical risks.

ECG monitorquality control testingpreventive maintenance

倪伟杰

展开 >

中国人民解放军陆军第七十三集团军医院(福建 厦门 361000)

心电监护仪 质量控制检测 预防性维护

2024

中国医疗器械信息
中国医疗器械行业协会

中国医疗器械信息

影响因子:0.375
ISSN:1006-6586
年,卷(期):2024.30(8)
  • 7