Comparison Between Real-Time Blood Glucose Testing Instrument and Beckman AU5800 Fully Automatic Biochemical Analyzer for Detecting Blood Glucoseevaluation of Indicator Results
Objective:To compare and evaluate the results of real-time blood glucose testing instrument and Beckman AU5800 fully automatic biochemical analyzer in detecting blood glucose indicators.Methods:150 patients with diabetes who underwent blood glucose examination in our hospital from August 2021 to August 2022 were selected for research.All patients were tested by real-time blood glucose tester and automatic biochemical analyzer,and the blood glucose results detected by different instruments,the precision of different blood glucose testers,and the correlation between the two blood glucose testers were analyzed.Results:There was no significant difference in the blood glucose detection results between the two blood glucose meters,P>0.05;The coefficient of variation of both blood glucose meters is less than 5%,indicating that the precision of both blood glucose detectors is relatively ideal;Regression analysis was conducted using a fully automated biochemical analyzer as the dependent variable and a real-time blood glucose meter as the independent variable.The results showed a good correlation between the two blood glucose meters.Conclusion:Both instruments have certain advantages in detecting blood sugar,and blood sugar detection methods should be determined based on the specific situation and needs of patients