Accuracy of CS-5100 Fully Automatic Coagulation Analyzer in Detecting Coagulation Dysfunction——A Case Study of Patients with Chronic Obstructive Pulmonary Disease
Objective:To analyze the accuracy of HISENMECON CS-5100 automatic hemagglutination instrument in detecting abnormal coagulation function.Methods:81 patients who were diagnosed by HISSMECOM CS-5100 automatic hemagglutination instrument from August 2020 to August 2023 were selected as the research objects,and the clinical test performance of HISSMECOM CS-5100 automatic hemagglutination.Results:The items tested this time include D-D(D dimer),Fib(fibrinogen),FBG(fasting blood glucose),APTT(activated partial thromboplastin time),PT(prothrombin time)and TT(thrombin time).In the inter-batch and intra-batch precision verification tests,the instrument met the manufacturer's reference range standards,and the CV was all less than 10.00%.The manufacturer's carrying pollution rate was≤10.00%,which meets the manufacturer's fixed value standard and can pass the verification.Conclusion:The application of HISENMECON CS-5100 automatic hemagglutination instrument has a good precision in detecting APTT,TT,PT and other coagulation indexes,which can provide a reference for doctors to evaluate the effect of diagnosis and treatment of disease and prognosis recovery.The instrument has ideal inspection precision and can bring accurate results for index detection.