首页|宽波束谐波运动成像定位高强度聚焦超声治疗焦点:体外研究

宽波束谐波运动成像定位高强度聚焦超声治疗焦点:体外研究

扫码查看
目的 观察超声宽波束(WB)谐波运动成像(HMI)(WB-HMI)用于定位离体组织高强度聚焦超声(HIFU)辐照焦点的有效性.方法 开发以声辐射力和超声成像为关键的WB-HMI技术.针对离体猪里脊和牛肝组织,采用不同幅度调制(AM)频率(25~100 Hz)和激励声功率(0.7~28 W)实现WB-HMI、定位HIFU辐照焦点,观察不同参数组合下WB-HMI定位辐照焦点的差异;以HIFU后声像图中的实际焦点位置为标准,以焦点定位误差<1 mm为WB-HMI定位焦点有效的标准,计算定位成功率.结果 相同AM频率下,声功率越大,WB-HMI定位焦点的位移幅值越大;相同声功率激励下,AM频率越小,WB-HMI定位焦点的位移幅值越大.对于离体猪里脊组织,不同AM频率下,声功率为15 W或22 W时,WB-HMI定位HIFU辐射焦点成功率均高于90.00%,而声功率为28 W时,定位成功率呈降低趋势;对于离体牛肝组织,声功率≥7.0 W时,WB-HMI定位成功率达100%.结论 WB-HMI能有效定位HIFU辐照离体组织焦点.
Wide beam-harmonic motion imaging for locating focus of high intensity focused ultrasound:An in vitro study
Objective To observe the effectiveness of ultrasound wide beam(WB)-based harmonic motion imaging(HMI)(WB-HMI)for locating irradiation focus of high intensity focused ultrasound(HIFU)for in vitro tissue.Methods WB-HMI technology was developed with acoustic radiation force and ultrasound imaging as the key technology.For in vitro porcine tenderloin and bovine liver tissue,different amplitude modulation(AM)frequencies(25-100 Hz)and excitation acoustic power(0.7-28 W)were used to achieve WB-HMI localization of HIFU irradiation focus,and the differences of WB-HMI localization of HIFU irradiation focus under different parameter combinations were observed.Taken the actual focus position on ultrasonic image after HIFU as the standard and the focus positioning error<1 mm as the effective standard of WB-HMI locating irradiation focus,the locating success rate was calculated.Results The larger the acoustic power,the larger the displacement amplitude of irradiation focus by WB-HMI at the same AM frequency,while the smaller the AM frequency,the larger the displacement amplitude of irradiation focus located by WB-HMI under the same acoustic power.Under different AM frequencies,for in vitro porcine tenderloin,the success rate of WB-HMI for locating HIFU radiation focus was higher than 90.00%when acoustic power was 15 W or 22 W,whereas the success rate showed a decreasing trend when the acoustic power was 28 W.For in vitro bovine liver tissue,the success rate of WB-HMI localization was 100%when acoustic power was ≥7.0 W.Conclusion WB-HMI could be used to effectively locate HIFU irradiation focus for isolated tissue.

extracorporeal shockwave therapyharmonic motion imagingwide beam imaging

冉瑶、徐佳虹、史新旺、刘依静、蔡德佳、周小伟

展开 >

重庆医科大学生物医学工程学院,超声医学工程国家重点实验室,重庆 400016

体外冲击波疗法 谐波运动成像 宽波束成像

2024

中国医学影像技术
中国科学院声学研究所

中国医学影像技术

CSTPCD北大核心
影响因子:0.763
ISSN:1003-3289
年,卷(期):2024.40(11)