首页|佳能CR-2型眼底照相机故障分析及解决方案

佳能CR-2型眼底照相机故障分析及解决方案

扫码查看
介绍了佳能CR-2型眼底照相机的工作原理和结构特点,分析了导致眼底照相机报故障码Err20、无法开机、图像有阴影等故障的原因,提出了快速有效的检修方法及处理方案,供同行参考和借鉴.
Failure Analysis and Solution of Canon CR-2 Digital Retinal Camera
It introduces the working principle and structural characteristics of the Canon CR-2 digital retinal camera.It analyzes the reasons that cause the fundus camera to report fault code Err20,cannot be turned on,and has shadows in the image.It also proposes quick and effective maintenance methods and solutions,for reference by peers.

Fundus cameraEquipment maintenanceFault analysisPreventive maintenance

熊伯通、李伟东

展开 >

广州医科大学附属第六医院,清远,511518

眼底照相机 设备维修 故障分析 预防性维护

2024

中国医院建筑与装备
卫生部医院管理研究所

中国医院建筑与装备

影响因子:0.304
ISSN:1671-9174
年,卷(期):2024.25(1)
  • 12