肿瘤学杂志2024,Vol.30Issue(6) :498-503.DOI:10.11735/j.issn.1671-170X.2024.06.B008

同发双侧乳腺癌保乳术后患者螺旋断层调强放疗与容积弧形调强放疗的剂量学对比分析

Dosimetric Comparison Between Helical Tomotherapy and Volumetric Modulated Arc Therapy for Patients with Synchronous Bilateral Breast Cancer After Breast-Conserving Surgery

刘楠 董婷婷 杨波 庞廷田 张杰 周震 张文君 于浪
肿瘤学杂志2024,Vol.30Issue(6) :498-503.DOI:10.11735/j.issn.1671-170X.2024.06.B008

同发双侧乳腺癌保乳术后患者螺旋断层调强放疗与容积弧形调强放疗的剂量学对比分析

Dosimetric Comparison Between Helical Tomotherapy and Volumetric Modulated Arc Therapy for Patients with Synchronous Bilateral Breast Cancer After Breast-Conserving Surgery

刘楠 1董婷婷 1杨波 1庞廷田 1张杰 1周震 1张文君 1于浪1
扫码查看

作者信息

  • 1. 中国医学科学院北京协和医院,北京 100730
  • 折叠

摘要

[目的]比较同发双侧乳腺癌保乳术后患者采用螺旋断层调强放疗(helical tomotherapy,TOMO)技术相较于容积弧形调强放射治疗(volumetric modulated arc therapy,VMAT)技术的剂量学差异.[方法]选取既往采用VMAT治疗的同发双侧乳腺癌保乳术后患者10例,重新进行TOMO计划设计,分别评价靶区的D2%、D98%、平均剂量Dmean以及均匀性指数(homogeneity index,HI)、适形度指数(comformal index,CI)和梯度指数(gradient index,GI),危及器官包括心脏、左冠状动脉前降支、双肺、左肺、右肺、食管、肝脏等,并统计机器跳数、出束时间.[结果]TOMO计划的靶区Dmean相对较低且高剂量D2%低于VMAT计划(P=0.004、0.002),低剂量D98%两组相当(P=0.444).在靶区的CI和GI方面,TOMO计划更具优势(P均<0.001),两组HI差异无统计学意义(P=0.120).TOMO计划的心脏V5Gy、Dmean和左冠状动脉前降支的Dmax高于VMAT计划,差异均有统计学意义(P<0.001,P=0.007,P=0.031);TOMO计划双肺VaGy高于VMAT计划(P=0.048);TOMO计划的肝脏Dmean高于VMAT计划(P<0.001).与VMAT相比,TOMO计划机器跳数增加明显且计划执行时间增加约4.5倍.[结论]在同发双侧乳腺癌中,TOMO与VMAT相比在靶区包绕适形性和剂量跌落上优于常规VMAT技术,除食管外未发现危及器官剂量学优势,计划执行效率上有待提高.

Abstract

[Objective]To compare the dosimetry difference between helical tomotherapy(TOMO)and volu-metric modulated arc therapy(VMAT)plan in patients with synchronous bilateral breast cancer(SBBC)after breast-conserving surgery.[Methods]Ten patients with SBBC who underwent VMAT after breast-conserving surgery were enrolled in the study.The TOMO plan was re-designed;and D98%,Dmean,conformation index(CI),homogeneity index(HI)and gradient index(GI)of the target area were evaluated.The receiving capacity of the organs at risk including heart,left anterior descending coronary artery,both lungs,left lung,right lung,esophagus and liver were calculated,and monitor units(MU)and beam-on time(BOT)were counted.[Results]The Dmean of the target area of TOMO plan and the high-dose D2% were lower than those of VMAT plan(P=0.004,0.002),while the low-dose D98% was not significantly different between two groups(P=0.444).In terms of CI and GI,TOMO was significantly more advantageous(all P<0.001),and there is no significant difference in HI between the two groups(P=0.120).The V5 Gy,Dmean,and Dmax of the heart and left anterior descending coronary artery in TOMO plan were significantly higher than those in VMAT plan(P<0.001,P=0.007 and P=0.031).The V20 Gy of both lungs in TOMO plan was higher than that in VMAT plan(P=0.048).The Dmean of liver in TOMO plan was higher than that of VMAT plan(P<0.001).Compared with VMAT,MU of TOMO plan in-creased significantly and the BOT increases about 4.5 times.[Conclusion]For SBBC patients,TOMO is supe-rior to VMAT in terms of target conformation and dose fall off.No dosimetric advantage of the organs at risk was found except in the esophagus,and the efficiency of the plan execution needs to be improved.

关键词

乳腺肿瘤/同发双侧/螺旋断层调强放疗/容积弧形调强放射治疗/剂量学比较

Key words

breast neoplasms/simultaneous bilateral/helical tomotherapy/volumetric modulated arc therapy/dosimetric comparition

引用本文复制引用

基金项目

中央高水平医院临床科研专项(2022-PUMCH-A-101)

出版年

2024
肿瘤学杂志
浙江省肿瘤医院 浙江省抗癌协会

肿瘤学杂志

CSTPCD
影响因子:0.83
ISSN:1671-170X
段落导航相关论文