首页|New Co-Cr-Mo-Nb-Cu alloy for implant applications: Properties characterization

New Co-Cr-Mo-Nb-Cu alloy for implant applications: Properties characterization

扫码查看
? 2022 Elsevier B.V.In this research, a novel antibacterial alloy was fabricated through the addition of Cu and Nb to the conventional Co-Cr-Mo (CCM) alloy. Tungsten inert gas (TIG) welding was utilized to add Cu and Nb to the CCM alloy by surface re-melting. The microstructural investigations revealed that the addition of Cu and Nb to the CCM alloy resulted in significant changes. It was also concluded that the addition of Nb up to the optimum value of 0.2 wt%Nb increased the wear resistance, however, no significant effect was observed on corrosion resistance. Moreover, it has been found that the optimum alloy showed higher charge transfer resistance (Rct) in SBF solution and also better protein adsorption. In addition, it was concluded that the modification of the CCM alloy with 4.5%wt Cu and 0.2 wt%Nb led to a decrease in the Co concentration released in SBF solution, and boosted the antibacterial property of the CCM alloy.

Co-Cr-Mo alloyCorrosionProtein adsorptionSurface re-meltingWear

Mohammad Moradi S.、Zangeneh S.、Vardak S.、Bahrami R.

展开 >

Department of Materials and Textile Engineering Faculty of Engineering Razi University

Department of Materials Engineering Isfahan University of Technology

Department of Materials and Metallurgical Engineering Amirkabir University of Technology (Tehran Polytechnic)

2022

Journal of Alloys and Compounds

Journal of Alloys and Compounds

EISCI
ISSN:0925-8388
年,卷(期):2022.925
  • 2